Horizontal Steam Sterilizer, STH-NS Series
설명
Application
The horizontal steam sterilizer utilizes saturated steam as the sterilization medium. It leverages the physical property of saturated steam releasing a large amount of latent heat during condensation to penetrate items and achieve sterilization through a period of thermal insulation. The equipment features fully automatic control via PLC and HMI, including automatic temperature recording and vacuum drying functions. It is designed for the sterilization and drying of medical textiles, surgical instruments and apparatus, culture media, and glassware. It is widely used for disinfection and sterilization in hospital supply centers and operating rooms, pharmaceutical factories, laboratories, and research institutions.
Features
- Horizontal rectangular double-layer structure for excellent safety, heat preservation, and drying performance.
- Gear-engagement + motorized door locking structure, with low failure rate and high reliability.
- Pneumatic door sealing; Optional single or double doors (with double-door interlocking).
- Equipped with standard equipment validation interfaces.
- Optimized piping system with high-quality stainless steel tubes and clamp-type quick-release fittings.
- Control system: HMI with touch screen + PLC.
- Dual assurance of sterilization effect via F0 value and temperature-time monitoring, with comprehensive sterilization cycle records.
- Three-level authority management.
- Multiple safety protection measures.
- Pulsing Cycles: 0~9 times (adjustable).
- Sterilization programs: Fabric, instruments, liquid, BD test, pressure hold test, and multiple custom programs.
Main Technical Parameters
| Design Pressure | -0.1~0.3MPa | Temperature Uniformity | ≤±2.0℃ |
| Design Temperature | 150℃ | Water Source Pressure | 0.1~0.2MPa |
| Working Pressure | 0.23MPa | Compressed Air Pressure | 0.4~0.7MPa |
| Working Temperature | 115~138℃ | External Steam Pressure | 0.3~0.6MPa |
| Vacuum Level | -0.086MPa | Power Supply | 3-phase 5-wire, AC 380V/50Hz |
Specifications
| Model | Capacity (L) | External Dimensions (L*W*H) (mm) | Chamber Dimensions (L*W*H) (mm) |
Steam Consumption (kg) |
Water Consumption (kg) |
Power (kW) |
Gross Weight (Single/Double Door) (kg) | Net Weight (Single/Double Door) (kg) |
|
STH-NS150DM/ STH-NS150SM |
150 | 950x1120x1760 | 700x450x480 | ≤14 | ≤190 | ≤23 | 820/850 | 720/750 |
|
STH-NS200DM/ STH-NS200SM |
200 | 1175x1120x1760 | 925x450x480 | ≤16 | ≤190 | ≤23 | 890/920 | 770/800 |
|
STH-NS250DM/ STH-NS250SM |
250 | 1030x1300x1900 | 780x600x600 | ≤18 | ≤200 | ≤35 | 990/1020 | 840/870 |
|
STH-NS360DM/ STH-NS360SM |
360 | 1250x1300x1900 | 1000x600x600 | ≤25 | ≤200 | ≤35 | 1090/1130 | 910/950 |
|
STH-NS600DM/ STH-NS600SM |
600 | 1440x1310x1900 | 1190x610x910 | ≤30 | ≤210 | ≤50 | 1440/1490 | 1240/1290 |
|
STH-NS800DM/ STH-NS800SM |
800 | 1700x1310x1900 | 1450x610x910 | ≤35 | ≤200 | ≤5.0 | 1590/1650 | 1405/1465 |
|
STH-NS1000DM/ STH-NS1000SM |
1000 | 1500x1440x1950 | 1250x680x1180 | ≤42 | ≤200 | ≤5.0 | 1680/1730 | 1550/1610 |
|
STH-NS1200DM/ STH-NS1200SM |
1200 | 1700x1440x1950 | 1450x680x1180 | ≤50 | ≤200 | ≤6.0 | 1870/1950 | 1650/1730 |
|
STH-NS1500DM/ STH-NS1500SM |
1500 | 2100x1440x1950 | 1850x680x1180 | ≤60 | ≤200 | ≤6.0 | 2180/2250 | 1900/1975 |
Note:
- Models 150L-600L are standard-equipped with a built-in steam generator; models above 600L require an external steam source.
- Sterilizers below 600L include sterilization racks as standard; models 600L and above include sterilization carts and transfer trolleys as standard.
- In the model numbers, “D” denotes Single Door, “S” denotes Double Door, and “M” denotes Pulse Vacuum.